    Abstract:

    To develop a chemiluminescent immunoassay method for detecting serum ferritin with biotin-avidin amplified system and a sandwich way. A pair of antibodies were labeled with biotin and horseradish peroxidase respectively. The 96 well microplate was coated with streptavidin. The serum ferritin standard and international standard was evaluated. 300 clinical serum specimens wer detected with this method and a foreign serum ferritin kit(Roche),the correlation coefficient was 0.97. The linear measurement range of this method was 10~800 ng/mL. The intra-assay and inter-assay coefficients of variation were 3.4~5.5%, and 5.0~7.3%. The cross-reactivity rate with CEA(carcinoembryonic antigen),human albumin,human heart ferritin,and hemoglobin were all less than 1%. The developed method was stable 9 days at 37 ℃. This assay was easily operable,small amount of specimens needed and the reaction was quick,so it was suitable for clinical application.

    Reference
    Related
    Cited by
Get Citation

SUN Shanshan, LI Xin, SONG Cuicui, YANG Yankun, BAI Zhonghu. Construction and Clinical Application of Chemiluminescent Immunoassay for Human Serum Ferritin[J]. Journal of Food Science and Biotechnology,2019,38(3):118-123.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:
  • Adopted:
  • Online: June 21,2019
  • Published:
Article QR Code

Copy Right:Editorial Board of Journal of Food Science and Biotechnology

Address:No. 1800, Lihu Avenue, Wuxi 214122, Jiangsu Province,China  PostCode:214122

Phone:0510-85913526  E-mail:xbbjb@jiangnan.edu.cn

Supported by:Beijing E-Tiller Technology Development Co., Ltd.

WeChat

Mobile website